Product Specification
| Brand | ADEX International |
| Condition | New |
| Warranty | YES |
Product Descriptions
Product Category: Measuring Instruments
Brand: HIOKI
PRODUCT OVERVIEW
| DC Voltage | 400.0 mV to 1000 V, 5 ranges, Basic accuracy: ±0.5 % rdg. ±3 dgt. |
|---|---|
| AC Voltage | 400.0 mV to 1000 V, 5 ranges, Frequency characteristics: 40 Hz - 500 Hz Basic accuracy 40 - 500 Hz : ±1.0 % rdg. ±5 dgt. (Average rectified) |
| Resistance | 400.0 Ω to 40.00 MΩ, 6 ranges, Basic accuracy: ±0.5 % rdg. ±2 dgt. |
| DC Current | 400.0 μA to 10.00 A, 6 ranges, Basic accuracy: ±1.2 % rdg. ±3 dgt. |
| AC Current | 400.0 μV to 10.00 A, 6 ranges, Frequency characteristics: 40 Hz - 500 Hz Basic accuracy 40 - 500 Hz : ±1.2 % rdg. ±5 dgt. (Average rectified) |
| Temperature (use with DT4910) | N/A |
| Voltage detection | N/A |
| C (capacitance) | 50.00 nF to 100.0 μF, 5 ranges, Basic accuracy: ±2.0 % rdg. ±5 dgt. |
| Frequency | AC V measurement: 5.000 Hz (1 Hz or more) to 5.000 MHz, 7 ranges, Basic accuracy: ±0.1 % rdg. ±3 dgt., Minimum sensitivity voltage: Square wave of 1.5 Vrms |
| Continuity check | Continuity threshold [ON]: 90 Ω ±40 Ω or less (Buzzer sound) |
| Diode test | Open terminal voltage: 3.0 V or less, Testing current 0.5 mA |
| Environment | Operating temperature: -10 to 50 °C, Storage temperature: -20 to 60 °C |
| Other functions | Display value hold, Relative display, Auto-power off, Back light |
| Display | 4-digits LCD, max. 4000 counts |
| Display refresh rates | 3 times/s |
| Power supply | LR6 alkaline batteries ×2, or R6P batteries ×2, Continuous use: 800 hr (When using LR6) |
| Dimensions and mass | 91.6 mm (3.61 in)W × 180.6 mm (7.11 in)H× 57.1 mm (2.25 in)D, 388 g (13.7 oz) (including batteries and holster) |
| Accessories | Test Lead L9206 ×1, holster (attached) ×1, instruction manual ×1, R6P manganese battery ×2 |
Safe and secure payments using Abraa safe trade systems
\